• AIRFLOW PUF HVS MICRO POLLUTANT HIGH VOLUME SAMPLER

Reference methods: Dir.CEE 107-2004 - ISO 12884 (Gas and Particles) ISO 16362 (Particles) - US-EPA TO4A, TO9A e TO13A (Gas e Particles)US-EPA TO4A, TO9A e TO13A (Gas and Particles) - ASTM D-6209

SAMPLING FEATURES

The Air Flow PUF high volume sampler with the new micro processor allow the user to view alL setting and sampling data for full compliance sampling status. The following data are available at display in a different screens:

  • • Selected sampling time (Max 9999 minute)
  • • Residual sampling time to stop
  • • Ambient Temperature
  • • Sampler cabinet temperature
  • • Barometric atmospheric pressure
  • • Sampling flow-rate
  • • Total sampling volume
  • • Total standardized sampling volume
  • • Sampling faults (flow or sensors readings)
  • • Wind velocity and direction, rain and humidity (with optional sensors installed)
  • • Stored sampling data to download with RS232 or USB port

The new microprocessor capabilities allow the user to select and store in the memory up to 3 different programs with sampling settings ready to start. Four different parameters are available on the main screen in a graphic format as Ambient Temperature, Barometric Pressure, Flow rate and Standard Deviation. Automatic re-star t after power shut-off with all data stored in the memory.

The AirFlow PUF high volume sampler is a dedicated sampling system to collect simultaneously gas and particles in a flow rate range to

120-600 liters per minute. Fully in compliance with all standard methods EPA,ISO, ASTM, the AirFiow Puf high volume sampler is microprocessor controlled and a Venturi flow measurement system (ISO 5167 Compliance) keep the flow rate constant with a great accuracy.


Technical Specifications
External sampler shelter Non-corrosive anodised aluminum

Insulation Class IP55
Sampling Flow
Filter type
Flow control
Suction pump
Power
Absorption (Ampere)
Dimensions
100-1300 lit/minute (depending on size selector)
102mm circular
Electronic with Venturi ISO 5167 device
Blowing in three magnetic induction stages
220 Volt/50 Hz
15 at start / 4 during operating phases
120x37x37 cm (H x L x W)
Weight 18.0 Kg
CE Compliance For electric and electromagnetic safety



Download Brochure for further details.

AIRFLOW PUF HVS MICRO POLLUTANT HIGH VOLUME SAMPLER


Related Products

High Volume Air Sampler for PM10

High Volume Air Sampler for PM10

  FEATURES: For PM 10, Optional can be set on PM 2.5  US EPA & EU Compliance..

High Volume Air Sampler for PM2.5

High Volume Air Sampler for PM2.5

High volume sampler dedicated to PM2.5 particles with a suction flow of 500 litres/minute in complia..

Tags: AIRFLOW PUF HVS MICRO POLLUTANT HIGH VOLUME SAMPLER