US EPA Approved FEM Method for PM10
The Met One Instruments, Inc. E-BAM Plus Portable Beta Gauge has been built to satisfy users, regulators and those from the health community by providing truly accurate, precise, real time measurement of fine particulate matter automatically. In addition, it is rugged, portable, and deployable in 15 minutes. The EBAM Plus is a U.S.-EPA designated equivalent method for PM10
The E-BAM Plus Offers the Following Advanced Features:
• Accuracy and precision approval with U.S.-EPA designationfor PM10 measurement.
• Real-time, accurate results without correction factors, regardless of season or geographic location.
• True ambient sampling provides accurate measurement of semi-volatile nitrates and organic compounds.
• Lightweight, rugged construction is easily mounted on a tripod in minutes.
• All-weather construction allows for true ambient sampling.
• Operates on AC power.
The E-BAM Plus is a Complete Measurement System That Comes With the Following Standard Components and Features:
• Internal Data-logger
• External AC Vacuum Pump Standard
• Real-Time Concentration
• PM10 Inlet
• Aluminum Tripod
• Temp/RH/Pressure Sensor
• Volumetric Flow Control
• Weatherproof Enclosure
• Filter Temperature Sensor
• Filter RH Sensor
• Filter Pressure Sensor
• Calibration Membrane
The Met One Instruments, Inc. E-BAM Plus is a portable, real-time beta gauge is a U.S.-EPA designated equivalent method for PM10.
SPECIFICATIONS:
Measurement
Principle
Particulate Concentration by
Beta Attenuation
U.S. EPA
Designations Measurement
PM10 Designation Number: EQPM-1215-226
Measurement
Sample Time
1 Hour
Flow
Rate
16.7 L/min inlet flow rate;
actual volumetric flow
Filter
Tape
Continuous glass fiber filter;
30 mm x 21 m roll; > 60 days/roll
Span
Check
Manual, 800ug (typical), span
foil included.
Beta
Source
14C
(carbon-14); 60 µCi ±15 µCi (2.22 MBq)
Beta
Detector Type
Photomultiplier tube with
organic plastic scintillator
Operating
Temperature Range
-30° to +50°C.
Operating
Humidity Range
0 – 90% RH, noncondensing
Inlet
Humidity Control
Actively controlled inlet heater
module
User
Interface
4.3” graphical touch screen
Ambient
Sensor
Model 597A combination AT, RH,
and BP serial sensor AT: -50° to +70°C; RH: 0 to 98%; BP: 375 to 825 mmHg
Analog
Outputs
2 Channels; optically Isolated;
Voltage Range 0-1, 0-2.5 or 0-5 VDC, Current Ranges 4-20 MA
7500
Digital Serial Interface
2 channels, half duplex RS-485
Serial
Interface
1 channel; full duplex RS-232,
USB (Shared common serial output) 1 channel; half duplex RS-485, Baud rates
1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200
Internal
Data Storage
8 Days 1-minute average, 1.3
years 60-minute average
External
Data Storage
1 USB Flash drive device
Compatible
Software Power
Air Plus 5, COMET™,
HyperTerminal®
Supply
AC Version: 100 - 230 VAC; 50/60
Hz; 150W, 3A @ 115 VAC/ 2A @ 230 VAC
Weight
75 lbs. (34.1 kg)
Dimensions
(Without Tripod)
Height: 18 in (46 cm) Width: 16
in (41 cm) Depth: 12 in (31 cm)
Portable Automatic Continuous Real Time Particulate Monitor E-BAM Plus by Metone